CLINITEK STATUS+ ANALYZER CT STATUS+

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-08-25 for CLINITEK STATUS+ ANALYZER CT STATUS+ manufactured by Siemens Healthcare Diagnostics Mfg., Ltd..

Event Text Entries

[2178175] Pt was tested for hcg as pre-operation procedure. Pt sample was tested 4 times. Results alternated between positive and negative. No unnecessary medical procedure was performed. No treatment was withheld as a result of this incident. There was ni impact to pt health.
Patient Sequence No: 1, Text Type: D, B5


[9292123] Mfr requested instrument for eval. Initial test by mfr was a false positive result. Instrument inspected and debris observed on optic's mirror. Mirror cleaned and multiple repeat testing results were normal. No further problems were reported with the instrument.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2011-00031
MDR Report Key2240591
Report Source07
Date Received2011-08-25
Date of Report2011-08-02
Date of Event2011-08-02
Date Mfgr Received2011-08-02
Date Added to Maude2012-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK SMITH
Manufacturer Street2 EDGEWATER DR.
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693158
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MFG,. LTD.
Manufacturer StreetNORTHERN RD. SUDBURY
Manufacturer CitySUFFOLK CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINITEK STATUS+ ANALYZER
Generic NameCLINITEK STATUS+ ANALYZER
Product CodeLCX
Date Received2011-08-25
Model NumberCT STATUS+
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS MFG., LTD.
Manufacturer AddressNORTH ROAD SUDBURY SUFFOLK CO106DX UK CO10 6DX


Patients

Patient NumberTreatmentOutcomeDate
10 2011-08-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.