MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-08-25 for CLINITEK STATUS+ ANALYZER CT STATUS+ manufactured by Siemens Healthcare Diagnostics Mfg., Ltd..
[2178175]
Pt was tested for hcg as pre-operation procedure. Pt sample was tested 4 times. Results alternated between positive and negative. No unnecessary medical procedure was performed. No treatment was withheld as a result of this incident. There was ni impact to pt health.
Patient Sequence No: 1, Text Type: D, B5
[9292123]
Mfr requested instrument for eval. Initial test by mfr was a false positive result. Instrument inspected and debris observed on optic's mirror. Mirror cleaned and multiple repeat testing results were normal. No further problems were reported with the instrument.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1217157-2011-00031 |
MDR Report Key | 2240591 |
Report Source | 07 |
Date Received | 2011-08-25 |
Date of Report | 2011-08-02 |
Date of Event | 2011-08-02 |
Date Mfgr Received | 2011-08-02 |
Date Added to Maude | 2012-06-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MARK SMITH |
Manufacturer Street | 2 EDGEWATER DR. |
Manufacturer City | NORWOOD MA 02062 |
Manufacturer Country | US |
Manufacturer Postal | 02062 |
Manufacturer Phone | 7812693158 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS MFG,. LTD. |
Manufacturer Street | NORTHERN RD. SUDBURY |
Manufacturer City | SUFFOLK CO102XQ |
Manufacturer Country | UK |
Manufacturer Postal Code | CO10 2XQ |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINITEK STATUS+ ANALYZER |
Generic Name | CLINITEK STATUS+ ANALYZER |
Product Code | LCX |
Date Received | 2011-08-25 |
Model Number | CT STATUS+ |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS MFG., LTD. |
Manufacturer Address | NORTH ROAD SUDBURY SUFFOLK CO106DX UK CO10 6DX |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-08-25 |