CLINICAL CHEMISTRY ALKALINE PHOSPHATASE REAGENT 07D55-31

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2011-09-13 for CLINICAL CHEMISTRY ALKALINE PHOSPHATASE REAGENT 07D55-31 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[2383531] The customer observed high absorbance values for calibration of the clinical chemistry alkaline phosphatase assay since the first use of reagent lot 08145un11 on (b)(6) 2011. The customer stated quality controls were unaffected and no calibration failures or patient exceptions were observed. The customer did observe block spots inside the alkaline phosphatase reagent cartridges and performed microscopy. The customer determined the reagent's dark color was the result of fungal elements present in the reagent and requested replacement reagents. There was no impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


[9392049] No consequences or impact to patient. Contamination during use. The cause of the clinical chemistry alkaline phosphatase reagents failed to calibrate/generate results was due to visible particulate or visible mold in the reagent cartridges. Abbott issued a product recall letter to all customers to inform them to discontinue use and to discard/destroy any suspect cartridges and order replacement reagents.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2011-00611
MDR Report Key2246990
Report Source01
Date Received2011-09-13
Date of Report2011-08-16
Date of Event2011-08-13
Date Mfgr Received2011-08-16
Device Manufacturer Date2011-03-24
Date Added to Maude2011-10-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1628664-8/29/11-001-R
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY ALKALINE PHOSPHATASE REAGENT
Product CodeCJE
Date Received2011-09-13
Catalog Number07D55-31
Lot Number08145UN11
Device Expiration Date2012-02-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-13