SERACLONE ANTI-K (KEL1) 808 090 100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-08-23 for SERACLONE ANTI-K (KEL1) 808 090 100 manufactured by Bio-rad Medical Diagnostics Gmbh.

Event Text Entries

[2157538] The customer reported false negative reactions of seraclone anti-k with a kell heterozygous cell from reagent red blood cells of a competitor. The customer had sent us the allegedly defective sample and the affected competitor's reagent red blood cells. Both were rested and reacted correctly positive. The allegedly defective sample was tested with further k+k+ red cells of donors and commercially available panel cells. All positive reactions were correct and unambiguous positive. We did not observe any false negative reactions. Testing of our quality control laboratory confirmed the correct function of the affected lot of seraclone anti-k. A review of the batch record documentation showed no irregularities which might have affected negatively the quality of the complained product.
Patient Sequence No: 1, Text Type: D, B5


[9393185] Stn# 125230.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610824-2011-00113
MDR Report Key2254818
Report Source05
Date Received2011-08-23
Date of Report2011-08-12
Date of Event2011-07-22
Date Mfgr Received2011-07-22
Date Added to Maude2012-05-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDR. MARC GORZELLIK
Manufacturer StreetINDUSTRIESTRASSE 1
Manufacturer CityDREIEICH 63303
Manufacturer CountryGM
Manufacturer Postal63303
Manufacturer Phone103801528
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSERACLONE ANTI-K (KEL1)
Generic NameSERACLONE ANTI-K
Product CodeLKJ
Date Received2011-08-23
Returned To Mfg2011-08-03
Catalog Number808 090 100
Lot Number7007130-01
Device Expiration Date2011-08-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD MEDICAL DIAGNOSTICS GMBH
Manufacturer AddressINDUSTRIESTRASSE 1 DREIEICH D-63303 GM D-63303


Patients

Patient NumberTreatmentOutcomeDate
10 2011-08-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.