SENIOR SOLUTIONS 2CH STIM PKG 2863

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-09-15 for SENIOR SOLUTIONS 2CH STIM PKG 2863 manufactured by Dj Orthopedics De Mexico, S.a. De C.v..

Event Text Entries

[2195986] Therapist applied ifc to pt using 2 inch square electrodes and intensity set to pt tolerance. Electrodes used for treatment had been previously used for approx 12 treatments. During ifc treatment, an electrode began to pull away and the therapist discovered area under electrode was red with 2 tiny blisters; skin under remaining three electrodes also had small blisters. Pt stated she did not have a burning sensation and treatment felt the same as when previously applied. Prior to electrode application, skin was cleaned with soap and water; it is unk if (b)(4) electrodes were used during treatment. Medical treatment was interrupted at 12 minutes into a 20 minute session. There was no report of medical intervention. Pt precautions reported as obesity, shortness of breath, and fall precautions. Device user manual (b)(4): states to prepare treatment area by cleaning with medical grade alcohol, describes electrodes in configurations of 2. 75 inch round or 1. 25 inch round, states electrodes are for one time use, to use (b)(4) electrodes whenever possible. Djo considers any report of burn a serious injury and reportable event. Device has not been received for eval.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616086-2011-00102
MDR Report Key2255707
Report Source06
Date Received2011-09-15
Date of Report2011-09-15
Date of Event2011-08-17
Date Added to Maude2011-09-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1430 DECISION ST
Manufacturer CityVISTA CA 92081
Manufacturer CountryUS
Manufacturer Postal92081
Manufacturer Phone7607271280
Manufacturer G1DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
Manufacturer Street20230 SUBMETROPOLI EL FLORIDO CARRETERA LIBRE TIJUANA TECATE
Manufacturer CityTIJUANA 22244
Manufacturer CountryMX
Manufacturer Postal Code22244
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSENIOR SOLUTIONS 2CH STIM PKG
Generic NameELECTROTHERAPY
Product CodeIMJ
Date Received2011-09-15
Model Number2863
Lot NumberELECTRODE: 030224
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
Manufacturer Address20230 SUBMETROPOLI EL FLORIDO CARRETERA LIBRE TIJUANA TECATE TIJUANA 22244 MX 22244


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-09-15

© 2026 FDA.report