ACTION PATCH 199488-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-09-23 for ACTION PATCH 199488-001 manufactured by Empi, Inc..

Event Text Entries

[2196559] It was reported that when the customer was removing the action patch from the release liner card the patch pulled away from the battery. The patch was not used.
Patient Sequence No: 1, Text Type: D, B5


[9346593] The reason that pieces of the electrode assembly were coming apart upon attempted removal is that the electrode will not easily release from the (b)(4) coated side. The electrode was placed on the incorrect side of the release liner. This is a consequence of an operator error during manufacturing. The clinic did not use the electrode. "this mdr is being reported because in some situations such as a malfunction could contribute to a serious injury if it were to recur. "
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1721293-2011-00009
MDR Report Key2261257
Report Source06
Date Received2011-09-23
Date of Report2011-09-15
Date of Event2011-08-23
Date Mfgr Received2011-08-23
Date Added to Maude2012-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS DEBBIE MIRANDA
Manufacturer Street205 HWY 22 EAST
Manufacturer CityCLEAR LAKE SD 57226
Manufacturer CountryUS
Manufacturer Postal57226
Manufacturer Phone6058476357
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACTION PATCH
Generic NameACTION PATCH
Product CodeEGJ
Date Received2011-09-23
Returned To Mfg2011-08-30
Model Number199488-001
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerEMPI, INC.
Manufacturer Address205 HWY 22 EAST CLEAR LAKE SD 57226 US 57226


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.