CHAMPION SERIES 54

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-09-16 for CHAMPION SERIES 54 manufactured by Champion.

Event Text Entries

[2199225] Champion hemodialysis chair model 54 series, serial number (b)(4), foot mechanism has sharp metal edges. Pt's leg was resting off footrest and pt's leg was found to be cut and bleeding against sharp metal edge. This incident resulted in small skin tear. Rn placed dressing over the wound. Biomedical technologist notified this model of champion hemodialysis chair does not have padding or cover over foot mechanism. Older champion chairs have plastic or padding over mechanism. Newest chairs have plastic or padding over metal foot mechanism.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5022297
MDR Report Key2261364
Date Received2011-09-16
Date of Report2011-09-16
Date of Event2011-09-05
Date Added to Maude2011-09-26
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCHAMPION
Generic NameHEMODIALYSIS CHAIR
Product CodeFIA
Date Received2011-09-16
Model NumberSERIES 54
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerCHAMPION
Manufacturer AddressELKHART IN US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-09-16