SYNCHRON CX3 DELTA 467501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-09-26 for SYNCHRON CX3 DELTA 467501 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2197524] On (b)(6) 2011 customer reported that after performing the two week maintenance for the creatinine 3 (cre3) cup on the cx3 delta instrument, the cre3 cup began to leak creatinine reagent when primed. Customer reseated the photodetector, but it did not fix the leak, and customer then requested service for the instrument. No injuries were reported and no erroneous results were generated from this issue. Field service engineer (fse) advised customer to adjust the cre3 cup and total protein 3 (tp3) cup voltages. Customer then calibrated the instrument and ran a quality control check. No further problems have been reported.
Patient Sequence No: 1, Text Type: D, B5


[9328994] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-05731
MDR Report Key2264711
Report Source01,05,06
Date Received2011-09-26
Date of Report2011-08-31
Date of Event2011-08-31
Date Mfgr Received2011-08-31
Device Manufacturer Date1996-01-24
Date Added to Maude2012-06-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 SOUTH KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 SOUTH KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNCHRON CX3 DELTA
Generic NameANALYZER, CHEMISTRY, MICRO
Product CodeJJF
Date Received2011-09-26
Model NumberCX3 DELTA
Catalog Number467501
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 SOUTH KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.