NATUS BILIBAND REGULAR SIZE PN 900642

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,08 report with the FDA on 2011-09-07 for NATUS BILIBAND REGULAR SIZE PN 900642 manufactured by Olympic Medical Incorporated.

Event Text Entries

[17407405] Infant being treated with phototherapy somehow removed the label from the natus biliband eye protector; the label was found by nicu personnel in the infant's hand. No injury occurred. However, there was the potential of a choking hazard.
Patient Sequence No: 1, Text Type: D, B5


[17817665] Although the device that was the subject of this complaint was not available for eval, natus medical evaluated product from the same lot ((b)(4)) as well as product from a lot built before ((b)(4)) and a lot built after ((b)(4)) after the suspect lot. In all cases, the labels were adhered to the bilibands. Pull tests were performed on these labels. None of the labels tested were removed without a reasonable amount of force. (b)(4) has distributed bilibands for eight years. Since this is the first report of an infant removing the label during use and since the issue could not be duplicated with testing, (b)(4) believes this is an isolated incident. Other than submitting this mdr, no further action is advocated at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2011-00005
MDR Report Key2267858
Report Source01,05,08
Date Received2011-09-07
Date of Report2011-09-07
Date of Event2011-08-11
Date Mfgr Received2011-08-11
Device Manufacturer Date2010-11-01
Date Added to Maude2012-06-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNATUS BILIBAND
Generic NameNEONATAL EYE PAD
Product CodeFOK
Date Received2011-09-07
Model NumberREGULAR SIZE
Catalog NumberPN 900642
Lot NumberF11101210R
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPIC MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-07