MULTIX TOP 00451450

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-09-20 for MULTIX TOP 00451450 manufactured by Siemens Ag, Medical Solutions.

Event Text Entries

[2260075] Siemens healthcare became aware of the reported event through a medwatch report (b)(4) filed by the customer. A f/u call was made to the customer. According to the biomedical engineering supervisor, gary guyton, an elderly pt was being prepped for a shoulder x-ray on the multix sys. The pt was leaning on the table when the table floated causing her to fall. The pt fell on the floor and broke her hip. She was hospitalized at the facility.
Patient Sequence No: 1, Text Type: D, B5


[9353731] According to the biomedical engineering supervisor, (b)(4), the unit power was removed from the sys possibly due to a power loss. The sys is equipped with a floating table top that may move when the unit looses power. The described event occurred due to the pt leaning on the table when the unit lost power. The reported issue appears to be a first occurrence and no similar issues have been reported from the field.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2240869-2011-00039
MDR Report Key2274711
Report Source06
Date Received2011-09-20
Date of Report2011-09-19
Date of Event2011-07-16
Date Mfgr Received2011-08-26
Date Added to Maude2011-10-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANASTASIA MASON
Manufacturer Street51 VALLEY STREAM PKWY. MS G01
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6102194834
Manufacturer G1SIEMENS AG, MEDICAL SOLUTIONS
Manufacturer StreetROENTGENSTRASSE 19-21
Manufacturer CityKEMNATH 95478
Manufacturer CountryGM
Manufacturer Postal Code95478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTIX TOP
Generic NameTABLE, RADIOGRAPHIC, NON-TILTING, POWER
Product CodeIZZ
Date Received2011-09-20
Model Number00451450
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS AG, MEDICAL SOLUTIONS
Manufacturer AddressROENTGENSTRASSE 19-21 KEMNATH 95478 GM 95478


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2011-09-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.