WET PREP TRAY WITH ESTEEM GLOVE 4465ANE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2011-10-05 for WET PREP TRAY WITH ESTEEM GLOVE 4465ANE manufactured by Carefusion.

Event Text Entries

[2346372] Complaint description from end user: "(b)(6), surgical coordinator, called? Writer on (b)(6) 2011.? She stated she was present during the episode, there was no harm to the patient.? The piece was easily removed during the procedure and the procedure went on without delay and with no residual effects anticipated.? The prep into the vaginal canal is quite deep in some patients so the prep stick is often seated deep inside the canal so as to swipe the cervix with bedatine.? The stick broke at the "neck. "? She said, it was very peculiar but no further events have occured. "
Patient Sequence No: 1, Text Type: D, B5


[9353783] (b)(4). No sample was provided by the customer; therefore an evaluation of the complaint device for deficiency of construction could not be performed. However samples of our current production process were reviewed and no problems related with this issue were observed. However, in similar complaints with this incident were identified the contour stick sponges were without a seal. Therefore, to confirm the issue the batch reported was evaluated for any issues related with this customer report. The product was manufactured, inspected and released in accordance with our internal procedures and no issues were observed on this complaint. (b)(4). A tension test is performed during our manufacturing process in order to verify the seal on the stick sponges. In addition, samples are taken from production every 2 hours to conduct a destructive pull test to detect any anomalies in the welding process. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030673-2011-00020
MDR Report Key2277601
Report Source08
Date Received2011-10-05
Date of Report2011-10-05
Date of Event2011-07-29
Date Mfgr Received2011-07-29
Date Added to Maude2011-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street1430 WAUKEGAN ROAD
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8474737294
Manufacturer G1PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V
Manufacturer StreetCERRADA VIA DE LA PRODUCCION NO85PARQUE INDUST.MEXICALI III
Manufacturer CityMEXICALI 21600
Manufacturer CountryMX
Manufacturer Postal Code21600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWET PREP TRAY WITH ESTEEM GLOVE
Generic NameAPPLICATOR, ABSORBENT TIPPED, STERILE
Product CodeOJU
Date Received2011-10-05
Model Number4465ANE
Lot Number0000317223
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address1500 WAUKEGAN ROAD MCGAW PARK IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.