[2349314] Returned power cords received on (b)(6) 2011 for reported poor design. No mention of sparking was made in reference to the returned power supplies at this time. Twenty-four additional power supplies were received in product support (b)(6) 2011. Customer reported on (b)(6) 2011 that they have a power cord that sparked due to fraying at the adaptor. Customer wraps the power cords around the meter when moving from screening site to screening site. Customer did not know which cord had caused the sparking.
Patient Sequence No: 1, Text Type: D, B5


[9310210] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2011-01980
MDR Report Key2281116
Report Source05,06
Date Received2011-09-15
Date of Report2011-09-16
Date of Event2011-04-11
Date Mfgr Received2011-08-17
Date Added to Maude2011-11-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MANAGER
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHOLESTECH
Generic NameCHOLESTEROL METER
Product CodeCHH
Date Received2011-09-15
Returned To Mfg2011-06-22
Model Number14-874
Lot NumberNG
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-15