LEICA RM2125

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-09-30 for LEICA RM2125 manufactured by Leica Biosystems, Gmbh..

Event Text Entries

[2347428] Customer reported that while cutting with the microtome, the universal cassette clamp did not hold the blocks tightly. The tissue samples were pulled out from the embedding material, however, they were not destroyed and could be further processed and were able to be diagnosed.
Patient Sequence No: 1, Text Type: D, B5


[9354827] The investigation indicated the cause for the malfunction was due to the incorrect functioning of the universal cassette clamp. The cassettes which were used were not properly held by the clamp. The analysis of the affected clamp has been shown that one of the two springs is broken. Thereby a stable terminal of the clamp is no longer possible.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010478-2011-00009
MDR Report Key2281602
Report Source05
Date Received2011-09-30
Date of Report2011-09-08
Date of Event2011-09-01
Date Mfgr Received2011-09-08
Device Manufacturer Date1999-02-01
Date Added to Maude2012-06-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEFFEN LAABS
Manufacturer StreetHEIDELBERGER STR. 17-19
Manufacturer CityNUSSLOCH 69226
Manufacturer Postal69226
Manufacturer Phone224143345
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA RM2125
Generic NameMICROTOME
Product CodeIDL
Date Received2011-09-30
Model NumberRM2125
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS, GMBH.
Manufacturer AddressHEIDELBERGER STR. 17-19 NUSSLOCH 69226 69226


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-30

© 2026 FDA.report