RPM RESPIRATORY GATING SYSTEM H50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-10-05 for RPM RESPIRATORY GATING SYSTEM H50 manufactured by Varian Medical Systems.

Event Text Entries

[16122216] The customer reports that while evaluating the alignrt sys (visionrt) with rpm, they found a deviation when producing a breathing motion for the phantom and monitoring the rpm amplitude with visionrt on and off. The phantom (a polystyrene mannequin) was set up and the gating box placed the base of the sternum, the tilt of the box was no more than 10 degrees from the camera plane. There was no report of serious injury as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[16333488] Although there was no reported injury in this case, the available info suggests a possible malfunction of the device. Though still under investigation, varian has determined that a mdr is appropriate as this possible malfunction, should it recur, could potentially result in serious injury. Add'l f/u to this mdr is expected upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916710-2011-00131
MDR Report Key2291600
Report Source01,05,06
Date Received2011-10-05
Date of Report2011-09-05
Date of Event2011-09-05
Date Mfgr Received2011-09-05
Device Manufacturer Date2009-11-01
Date Added to Maude2012-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHYSICIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL PIGNATARO, MGR
Manufacturer Street911 HANSEN WAY M/S C-255
Manufacturer CityPALO ALTO CA 94304
Manufacturer CountryUS
Manufacturer Postal94304
Manufacturer Phone6504246471
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRPM RESPIRATORY GATING SYSTEM
Generic NameACCELERATOR, LINEAR, MEDICAL
Product CodeIXO
Date Received2011-10-05
Model NumberH50
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVARIAN MEDICAL SYSTEMS
Manufacturer AddressPALO ALTO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.