LAPSAC TISSUE ENTRAPMENT POUCH J-LS-060401

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07,08 report with the FDA on 2011-10-14 for LAPSAC TISSUE ENTRAPMENT POUCH J-LS-060401 manufactured by Cook Urological Inc.

Event Text Entries

[19531214] On (b)(6) 2011, customer attempted to use the lapsac tissue entrapment pouch. During use, the sac tore releasing the contents. Surgery was delayed but there was no apparent harm to the pt noted at the time of the occurrence.
Patient Sequence No: 1, Text Type: D, B5


[19751116] One package containing a used lapsac (surgical tissue pouch) was received from the hospital. The lapsac has been ripped horizontally to the outer edges on one side of the pouch, approx 3cm from the bottom of the bag. The opposite side of the lapsac was ripped about half way across the pouch with a 1. 5cm section not torn noting the remaining section out to the edge of the pouch to be torn, also approx 3cm from the bottom of the pouch. In addition there were two vertical tears, both approx 2cm long, one of the vertical tears was l shape with a small portion of the material split and folded over on the outside of the pouch. One area of the lapsac has what appears to be a puncture with the material pushed outward in line with the ripped areas. This device has most likely been punctured by an instrument of undetermined origin. Two stock lapsacs were acquired to attempt to recreate the customers difficulty. Excessive pressure was applied to one side of the pouch noting no tearing occurred. Secondly, using a needle to puncture the center of one side of the pouch, then applied pressure to the puncture area with no tearing occurring. Excessive pressure was applied to the puncture area noting the pouch had torn as the used pouch returned. Additional info being requested without response at this time, should info become available, it will be forwarded.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1825146-2011-00042
MDR Report Key2301731
Report Source01,06,07,08
Date Received2011-10-14
Date of Report2011-09-16
Date of Event2011-08-10
Report Date2011-09-16
Date Mfgr Received2011-09-16
Device Manufacturer Date2011-05-11
Date Added to Maude2011-10-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRIS KILANDER
Manufacturer Street1100 W. MORGAN
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal47460
Manufacturer Phone8128294891
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLAPSAC TISSUE ENTRAPMENT POUCH
Generic NameKGY BAG, INTESTINE
Product CodeKGY
Date Received2011-10-14
Returned To Mfg2011-09-28
Model NumberNA
Catalog NumberJ-LS-060401
Lot NumberU2064600
Device Expiration Date2014-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK UROLOGICAL INC
Manufacturer AddressSPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.