TISSUE-TEK UNI-CASSETTE 4155-02 M7321-32P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-10-18 for TISSUE-TEK UNI-CASSETTE 4155-02 M7321-32P manufactured by Sakura Finetex Usa.

Event Text Entries

[2291553] A male with down syndrome and acute myelogenous leukemia (aml) was admitted for chemotherapy. A bone marrow biopsy was performed on day of the admission. The pathology department performed a microscopic examination of the specimen and then they sent the specimen to a reference laboratory for histologic processing. The reference laboratory called and informed they could not find the specimen. Something caused the cassette to open and they lost the specimen. The hospital and reference laboratories searched for the specimen, which was wrapped in "histo-wrap" paper. The specimen was not located. Evaluation of cassette determined a possible manufacturer defect. It appears the tab that secures the cassette in a closed position either broke off or was missing. Staff evaluated other cassettes received with the suspect cassette, and found another damaged cassette. No other reports of this problem have been submitted. I have a photograph of the cassette to submit to medsun. I also have a photo of the normal cassette next to a damaged one. We are not returning the damaged cassette to the manufacturer because laboratory personnel feel they have to have it on hand in case inspectors come to hospital and want to examine devices that were problematic.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2306122
MDR Report Key2306122
Date Received2011-10-18
Date of Report2011-10-18
Date of Event2011-08-03
Report Date2011-10-18
Date Reported to FDA2011-10-18
Date Added to Maude2011-10-25
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameTISSUE-TEK UNI-CASSETTE
Generic NameLAB SPECIMEN TISSUE-PROCESSING CONTAINER
Product CodeKER
Date Received2011-10-18
Model Number4155-02
Catalog NumberM7321-32P
Lot Number*
ID Number*
Device AvailabilityY
Device Age1 DY
Device Sequence No1
Device Event Key0
ManufacturerSAKURA FINETEX USA
Manufacturer Address1750 WEST 214TH STREET TORRANCE CA 90501 US 90501


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-18