OLYMPIC COOL-CAP SYSTEM 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-10-03 for OLYMPIC COOL-CAP SYSTEM 60010 manufactured by Natus Medical Incorporated.

Event Text Entries

[2218439] Infant being treated for (b)(6) with the olympic cool-cap system. Approx 14. 5 hours into treatment, a burning smell was noted. The infant was moved to a different isolette because the smell was thought to have been coming from the isolette. Approx 9. 5 hours later a strong burning smell was noted. A short while later three popping sound were heard and the device shut down. The infant was transported to another facility to complete the treatment.
Patient Sequence No: 1, Text Type: D, B5


[9370631] The cause of the device malfunction was a failed power supply. Four components within the power supply failed (a 4-pin connector, a power transistor, a resistor and diode). The power supply was purchased from (b)(4). This power supply issue is being investigated under (b)(4) that was opened by (b)(6) in (b)(6) 2010. The customer's device is being repaired and the failed power supply is being replaced with one that was purchased from (b)(6). This alternate power supply was approved by the fda for use with the cool-cap device on (b)(6) 2010 under pma supplement (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2011-00007
MDR Report Key2313555
Report Source05,06
Date Received2011-10-03
Date of Report2011-10-03
Date of Event2011-09-10
Date Mfgr Received2011-09-12
Device Manufacturer Date2008-03-01
Date Added to Maude2012-06-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP SYSTEM
Generic NameCOOLING CAP (INFANTS)
Product CodeMXM
Date Received2011-10-03
Returned To Mfg2011-09-25
Model Number60010
Catalog Number60010
Lot NumberNA
ID NumberVERSION 2.0 SOFTWARE
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-03