[16923816] It is standard to have a bowl of heparnized saline on the table for flushing catheters. Syringes are labeled with labels that are supplied in the pack. The labels began dissolving in the heparnized saline. During use the labels were sucked up into the syringe. The tech caught the issue before it was used on the patient. Sterile labels provided in packs are not impermeable ~ dissolved in heparinized saline flush.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2320701
MDR Report Key2320701
Date Received2011-10-20
Date of Report2011-09-30
Report Date2011-09-30
Date Reported to FDA2011-10-20
Date Added to Maude2011-11-02
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameEP STERILE PACK - LABEL SHEET
Generic NameLABEL SHEET
Product CodeLYV
Date Received2011-10-20
Model Number*
Catalog Number*
Lot Number11DB5727
ID Number*
OperatorOTHER
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES
Manufacturer Address1170 NORTHPOINT BLVD. PARK CITY IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-20