BIO-OSS COLLAGEN BLOCK 100 MG UNK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2007-07-24 for BIO-OSS COLLAGEN BLOCK 100 MG UNK manufactured by Geistlich Pharma Ag.

Event Text Entries

[21751211] Report by patient. (b)(6) 2004: implantation of bio-oss collagen for root resection. Pain and discharge from tooth after surgery. (b)(6) 2004: tooth extraction. Wound healing problems during 4 months afterwards. On (b)(6) 2005: removal of bio-oss collagen, normal healing after surgery. Since (b)(6) 2006: restart of pain in treated region. Damage of trigeminus nerve during surgery (exact time point or surgery not specified). Severe pain, treatment by special consultant for pain treatment since (b)(6) 2005 (medication: gabapentin, baclofen, amitryptiline). The patient provided medical records. The are written by hand and could therefore not deciphered completely. However, deviations from the patient's report could be identified (i. E. (b)(6) 2005: normal wound healing, whereas the patient described wound healing problems during that period).
Patient Sequence No: 1, Text Type: D, B5


[21925825] This event could be related to the patient's allergy to animal product as bio-os collagen contains porcine collagen. The labeling warns of the possibility of allergic reactions. As this appears to be labeled event, while serious, no further follow up is deemed necessary.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9614442-2006-00001
MDR Report Key2326197
Report Source04
Date Received2007-07-24
Date of Report2006-03-24
Date of Event2005-12-01
Date Mfgr Received2006-03-24
Device Manufacturer Date2003-10-31
Date Added to Maude2011-11-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBETTINA SCHMID
Manufacturer StreetBAHNHOFSTRASSE 40
Manufacturer CityWOLHUSEN 6110
Manufacturer CountrySZ
Manufacturer Postal6110
Manufacturer Phone14925678
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIO-OSS COLLAGEN
Generic NameDENTAL BONE VOID FILLER
Product CodeNPM
Date Received2007-07-24
Model NumberBLOCK 100 MG
Catalog NumberUNK
Lot Number030623
ID NumberNA
Device Expiration Date2007-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGEISTLICH PHARMA AG
Manufacturer AddressWOLHUSEN 6110 SZ 6110


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention; 2. Deathisabilit 2007-07-24

© 2026 FDA.report