IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-11-08 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[21751678] Customer called on (b)(4) 2011 reporting that the wash well of an immage 800 immunochemistry system was overflowing due to low vacuum errors. Customer noted that before obtaining low vacuum errors, qc was within established ranges and no erroneous results were generated. Field service engineer (fse) was dispatched on (b)(4) 2011 and rebuilt the vacuum pump and replaced the filters for the vacuum system.
Patient Sequence No: 1, Text Type: D, B5


[21926295] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-07353
MDR Report Key2327814
Report Source05,06
Date Received2011-11-08
Date of Report2011-10-13
Date of Event2011-10-13
Date Mfgr Received2011-10-13
Device Manufacturer Date2006-09-26
Date Added to Maude2012-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-11-08
Model NumberIMMAGE 800
Catalog NumberA15445
ID NumberSW VERSION 2.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.