ABBOTT AXSYM SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 1999-07-22 for ABBOTT AXSYM SYSTEM manufactured by Abbott Manufacturing Inc.

Event Text Entries

[140808] Customer experiencing erratic results with valproic acid. On 6/23/99 customer obtained a result of 0. 56ug/ml. Result was questioned by physician, sample was rerun with a result of 75. Oug/ml. Customer is not sure if tube contained fibrin. No report of impact to patient.
Patient Sequence No: 1, Text Type: D, B5


[6360802] Customer obtained a valproic acid result of 7. 9ug/ml, result was questioned by physician. Sample was rerun with a result of 66. 9ug/ml. Physician questioned result prior to treating patient. No report of injury.
Patient Sequence No: 2, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-1999-00039
MDR Report Key233291
Report Source06
Date Received1999-07-22
Date of Report1999-07-21
Date of Event1999-06-23
Date Mfgr Received1999-06-28
Device Manufacturer Date1997-10-01
Date Added to Maude1999-07-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABBOTT AXSYM SYSTEM
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeLCI
Date Received1999-07-22
Model NumberNA
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorUNKNOWN
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key226131
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DR. IRVING TX 75038 US
Baseline Brand NameABBOTT AXSYM SYSTEM
Baseline Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Baseline Model NoNA
Baseline Catalog No7A83-01
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 1999-07-22
20 1999-07-22

© 2026 FDA.report