PEDIATRIC SCALE 4802

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2011-08-12 for PEDIATRIC SCALE 4802 manufactured by Scale-tronix, Inc..

Event Text Entries

[20217750] The scale could easily be turned from kilograms to pounds. The software was supposed to be locked to prevent access to pounds. Kilograms are the only measure of weight to be used in the user's pediatric hospital. There was an opportunity that staff does not realize that the scale went back to pounds and record the weight ponds. Medication doses are based on mg/kg (milligrams per kilograms).
Patient Sequence No: 1, Text Type: D, B5


[20592448] Scales are normally configured to toggle between pounds and kilograms. Customer can request scale to be programmed at factory to kilograms only or it can be accomplished on-site using the instructions in the operation manual beginning on page 4. In this situation the scales (sns (b)(4)) were ordered from the hosp as a normally configured pediatric scale with pounds and kilograms active.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1419590-2011-00001
MDR Report Key2335177
Report Source99
Date Received2011-08-12
Date of Report2011-08-12
Date of Event2011-06-01
Date Mfgr Received2011-07-11
Device Manufacturer Date2010-10-01
Date Added to Maude2012-06-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street288 CARLTON DR
Manufacturer CityCAROL STREAM IL 60188
Manufacturer CountryUS
Manufacturer Postal60188
Manufacturer Phone6306353377
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePEDIATRIC SCALE
Generic NameSCALE
Product CodeFRW
Date Received2011-08-12
Model Number4802
Lot NumberUNK
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSCALE-TRONIX, INC.
Manufacturer Address288 CARLTON CAROL STREAM IL 60188 US 60188


Patients

Patient NumberTreatmentOutcomeDate
10 2011-08-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.