OLYMPUS AU5421 CLINICAL CHEMISTRY ANALYZER WITH 2 FLOWCELL ISE N1683100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-11-15 for OLYMPUS AU5421 CLINICAL CHEMISTRY ANALYZER WITH 2 FLOWCELL ISE N1683100 manufactured by Beckman Coulter Mishima K.k..

Event Text Entries

[2362607] A customer reported to beckman coulter, inc. (bec) that olympus au5421 clinical chemistry analyzer generated an erroneously elevated total protein result of 114 g/l. The result was not reported out of the laboratory as it was not consistent with the laboratory's expectation for the patient. Repeat testing produced 67 g/l, which was in line with the expected result. There was no effect to the patient.
Patient Sequence No: 1, Text Type: D, B5


[9464078] The sample was a serum collected in 7ml bd tube, and was centrifuged for 7 minutes at 3250 rpm, 20c. Qc produced expected results before and after the event. The laboratory assessed the reaction profile for this result and noted an unexpected electrical spike in early stages of the reaction curve. There was no error message or flagging associated with this event. Bec field service engineer (fse) visited the site next day and identified a blocked drainage from the cuvettes. The issue was corrected and the instrument has been performing acceptably.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-07619
MDR Report Key2337492
Report Source01,05,06
Date Received2011-11-15
Date of Report2011-10-18
Date of Event2011-10-17
Date Mfgr Received2011-10-18
Device Manufacturer Date2008-01-01
Date Added to Maude2012-07-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER MISHIMA K.K.
Manufacturer Street454-32 HIGASHINO,NAGAIZUMI-CHO SUNTO-GUN
Manufacturer CitySCHIZUOKA, JP-NOTA 411-0931
Manufacturer CountryJA
Manufacturer Postal Code411-0931
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS AU5421 CLINICAL CHEMISTRY ANALYZER WITH 2 FLOWCELL ISE
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeCEK
Date Received2011-11-15
Model NumberAU5421
Catalog NumberN1683100
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER MISHIMA K.K.
Manufacturer Address454-32 HIGASHINO,NAGAIZUMI-CHO SUNTO-GUN SCHIZUOKA, JP-NOTA 411-0931 JA 411-0931


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.