VITROS CHEMISTRY PRODUCTS OP REAGENT 6801997

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-11-16 for VITROS CHEMISTRY PRODUCTS OP REAGENT 6801997 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[17555757] The customer obtained a non-reproducible, negative vitros op result for a known positive quality control fluid processed on the vitros 5,1 fs chemistry system. Biased results of the magnitude and direction observed may lead to inappropriate physician action if the event were to occur undetected with a patient sample. No patient results were known to have been affected. There was no report of patient harm as a result of this event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[17817275] The investigation determined that a non-reproducible, false negative vitros op result was obtained for a known positive quality control fluid. Patient samples were not known to have been affected. The customer recalibrated their in-use vitros op reagent lot, and acceptable performance was obtained following recalibration. A definitive root cause for the event could not be determined. However, a control fluid issue and/or an equipment or reagent related issue cannot be ruled out as contributing factors.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319808-2011-00039
MDR Report Key2339413
Report Source05
Date Received2011-11-16
Date of Report2011-11-16
Date of Event2011-10-28
Date Mfgr Received2011-10-28
Device Manufacturer Date2011-06-07
Date Added to Maude2012-07-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street1000 LEE ROAD
Manufacturer CityROCHESTER NY 14606
Manufacturer CountryUS
Manufacturer Postal Code14606
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS OP REAGENT
Generic NameIN-VITRO DIAGNOSTIC
Product CodeDJG
Date Received2011-11-16
Catalog Number6801997
Lot Number1527-06-1506
Device Expiration Date2012-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.