ARCHITECT SIROLIMUS WHOLE BLOOD PRECIPITATION REAGENT 01L76-55

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-11-17 for ARCHITECT SIROLIMUS WHOLE BLOOD PRECIPITATION REAGENT 01L76-55 manufactured by Abbott Laboratories.

Event Text Entries

[2361793] A customer had bloodwork performed for an unspecified reason and discovered that liver parameters were elevated. The customer contacted abbott to inquire whether a splash to the arm during disposal of an unknown architect reagent approximately one month prior could have caused an increase in liver test results. The customer reported that their skin was intact but no information was provided regarding which reagent, volume of reagent in contact with skin, length of exposure time, first aid measures taken to address the skin exposure, nor whether there was any skin irritation or allergic response. The customer provided no information regarding any medical treatment or diagnostic testing other than results of blood tests. There was no liver function testing prior to the event for comparison. The bloodwork results for the patient were reviewed by the abbott medical director. The patient results indicate that the customer has elevated liver enzymes but the tests were negative for (b)(6). Abbott reviewed the list of reagents supplied to this customer and the associated material safety data sheets. A few of the reagents shipped to this customer were identified as being potentially toxic if they came into contact with skin. These reagents do have the potential to impact the liver and a pre-existing liver condition may be aggravated by exposure. The reagent package insert for these reagents clearly state to wear suitable protective clothing, gloves and eye/face protection when handling these products. At the current time, without additional information, it is not possible to ascertain whether these reagents were the subject of the exposure incident. No specific information about the type of treatment the customer received was provided.
Patient Sequence No: 1, Text Type: D, B5


[9484428] Liver dysfunction. Device handling issue in order to isolate the possible assays. We reviewed all architect assays that are utilized at the customer's facility and determined that architect cyclosporine whole blood precipitation reagent, list 1l75-55, architect sirolimus whole blood precipitation reagent, list 1l76-55 and architect tacrolimus whole blood precipitation reagent, list 1l77-55 all have the possibility of affecting the liver. Customer complaints for the three assays received to-date were reviewed to determine if others have experienced the same issue. Our review of this data did not identify any problematic complaint activity that would indicate this product is performing contrary to its claims. Additionally, the package inserts and material safety data sheets for the three assays were reviewed. This review did indicate that the assays do have the potential to affect the liver. However, safety precautions/instructions are provided to minimize the risk. The customer was referred to the precautions section of each assays package insert. Based on this investigation, we have concluded that the architect cyclosporine whole blood precipitation reagent, architect sirolimus whole blood precipitation reagent and architect tacrolimus whole blood precipitation reagent are performing as intended and meeting their safety, effectiveness and label claims.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2011-00690
MDR Report Key2341250
Report Source05
Date Received2011-11-17
Date of Report2011-10-31
Date of Event2011-05-27
Date Mfgr Received2011-10-31
Date Added to Maude2011-11-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 60064350
Manufacturer CountryUS
Manufacturer Postal Code60064 3500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT SIROLIMUS WHOLE BLOOD PRECIPITATION REAGENT
Product CodeNRP
Date Received2011-11-17
Catalog Number01L76-55
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-11-17

© 2024 FDA.report