[2352587] Customer called to report that the immage 800 immunochemistry system generated erroneous alpha-antitrypsin (aat) results. Customer also reported low quality control recovery. The root cause could not be determined; however, it appears as though this was a reagent-specific issue, and not an instrument issue, as the problems occurred once this reagent lot was used. A service request was not generated. Customer stated that although erroneous results were generated, patient treatment was not affected.
Patient Sequence No: 1, Text Type: D, B5


[9485002] The root cause of the issue was not determined, but it appears as though the issue was reagent-specific. (note: the beckman coulter, inc. Identifier for this report is (b)(4)). Reagent-specific issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-07625
MDR Report Key2342629
Report Source01,05,06
Date Received2011-11-18
Date of Report2011-09-25
Date of Event2011-09-25
Date Mfgr Received2011-09-25
Device Manufacturer Date2011-04-22
Date Added to Maude2011-11-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 928216232
Manufacturer CountryUS
Manufacturer Postal928216232
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821623
Manufacturer CountryUS
Manufacturer Postal Code92821 6232
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMAGE IMMUNOCHEMISTRY SYSTEM ALPHA-1-ANTITRYPSIN (AAT)
Generic NameIMMUNOCHEMICAL, CERULOPLASMIN
Product CodeDEM
Date Received2011-11-18
Model NumberIMMAGE AAT REAGENT
Catalog Number447740
Lot NumberM105003
Device Expiration Date2013-03-31
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821623 US 92821 6232


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-18