[2351774] After hooking up the ablation catheter and testing the ethicon gynecare thermachoice machine prior to using on the patient, the machine delivered a message, "error 5506. " a new catheter was tried, from the same lot, and the same message appeared after testing. The ethicon technician was called and stated that "error 5506" message indicated that there was an internal component issue and to take the machine out of service. This was done and the procedure was aborted. ======================manufacturer response for thermal balloon catheter, ghyecare thermachoice iii uterine balloon catheter (per site reporter). ======================no response. Feels issue is with the machine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2344268
MDR Report Key2344268
Date Received2011-11-11
Date of Report2011-11-11
Date of Event2011-11-09
Report Date2011-11-11
Date Reported to FDA2011-11-11
Date Added to Maude2011-11-21
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameETHICON GYNECARE
Generic NameTHERMAL ABLATION MACHINE
Product CodeMNB
Date Received2011-11-11
Model NumberTHERMACHOICE 2 EAS2000-1
Catalog Number*
Lot Number*
ID Number*
OperatorPHYSICIAN
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerETHICON, INC.
Manufacturer AddressROUTE 22 WEST SOMERVILLE NJ 08876015 US 08876 0151

Device Sequence Number: 2

Brand NameGHYECARE THERMACHOICE III UTERINE BALLOON CATHETER
Generic NameTHERMAL BALLOON CATHETER
Product CodeMKN
Date Received2011-11-11
Model Number*
Catalog NumberTC003
Lot NumberDBMG03
ID Number*
Device AvailabilityY
Device Age1 DY
Device Sequence No2
Device Event Key0
ManufacturerETHICON, INC.
Manufacturer AddressROUTE 22 WEST SOMERVILLE NJ 08876015 US 08876 0151


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-11