ACTH CHEMILUMINESCENCE IMMUNOASSAY KIT 60-4175

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1995-07-13 for ACTH CHEMILUMINESCENCE IMMUNOASSAY KIT 60-4175 manufactured by Nichols Inst..

Event Text Entries

[19242523] When correlating mfr's chemiluminesence immunoassay method for acth (x) against the irma immunoassay method for acth (y) rptr found a significant bias described by the equation y = 1. 279x + 6. 1. The mean result of 17 pt samples done by the irma method was 145. 8 pg/ml, the mean of the chemil method was 118. 7 pg/ml. The reference ranges reported by nichols for the two methods is almost identical. The irma range is 9-52 pg/ml and the chemiluminescence is 7-51 pg/ml. As established via phone calls, the mfr was aware of this bias and was in the process of restandardizing both methods and re-establishing the reference ranges. However, rptr was not notified of the bias which existed between the results of these 2 methods.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1006485
MDR Report Key23447
Date Received1995-07-13
Date of Report1995-06-28
Date of Event1995-06-16
Date Added to Maude1995-07-13
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameACTH CHEMILUMINESCENCE IMMUNOASSAY KIT
Generic NameIMMUNOASSAY KIT
Product CodeCKG
Date Received1995-07-13
Catalog Number60-4175
Lot Number501914
Device Expiration Date1996-01-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key23730
ManufacturerNICHOLS INST.
Manufacturer AddressSAN JUAN CAPISTRANO CA 92690 US


Patients

Patient NumberTreatmentOutcomeDate
10 1995-07-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.