BECKMAN TOTAL BILIRUBIN TRIGGERED KIT 476826

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2011-11-23 for BECKMAN TOTAL BILIRUBIN TRIGGERED KIT 476826 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2355354] The customer reported that they observed a leak coming from a synchron cx4/cx5 clinical chemistry analyzer reagent. No patient results were involved in this event. It is unknown as to whether the personnel involved in this event was wearing personal protective equipment at the time of occurrence however no personnel sought medical attention in association with this event. There was no chemical exposure to personnel uncovered wounds or mucous membranes. No death, injury or modification to patient treatment was associated with this event. No exposure control plan was in place at the facility. The material safety data sheet was not reviewed.
Patient Sequence No: 1, Text Type: D, B5


[9466984] Beckman coulter inc. Assessment of the supplied photos indicated that the cartridges were leaking from the bottom seam. This may be indicative of a cartridge manufacturing issue. No service was dispatched to the site for this event. A definitive root cause for the leak is not known at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-07858
MDR Report Key2346698
Report Source01,06
Date Received2011-11-23
Date of Report2011-11-09
Date of Event2011-10-26
Date Mfgr Received2011-11-09
Device Manufacturer Date2011-04-12
Date Added to Maude2012-07-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBECKMAN TOTAL BILIRUBIN TRIGGERED KIT
Generic NameNADH OXIDATION/NAD REDUCTION, ALT/SGPT
Product CodeCKA
Date Received2011-11-23
Model NumberNA
Catalog Number476826
Lot NumberZ104209
ID NumberNA
OperatorOTHER
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BOULEVARD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-23