OLYMPUS VIDEO TELESCOPES WA50023B UNK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2011-11-17 for OLYMPUS VIDEO TELESCOPES WA50023B UNK manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[21759987] Olympus was informed that during a therapeutic laparoscopic cholecystectomy procedure, the image started out darker than usual, and lines began to develop on the image approx 5-10 mins into the procedure, the image was said to have been progressively getting worst which ultimately became invisible. The intended procedure was said to have been completed with a laparoscope. There was no pt harm reported.
Patient Sequence No: 1, Text Type: D, B5


[21869306] The device referenced in this report was returned to olympus for eval. The eval confirmed the user's report. The eval found the image flickered with horizontal lines and blackout completely after being burn-in for 10 mins. The outer tube was bent with multiple dents noted around the frame at the distal tip. The reported phenomenon was attributed to the light guide cable unit. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2011-00034
MDR Report Key2351003
Report Source06,07
Date Received2011-11-17
Date of Report2011-10-18
Date Mfgr Received2011-10-18
Date Added to Maude2012-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPES
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2011-11-17
Returned To Mfg2011-10-21
Model NumberWA50023B
Catalog NumberUNK
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-17