[2274282] The user facility reported that during an unspecified procedure, the users experienced a complete loss of image. The procedure was completed using a different device. There was no further information provided.
Patient Sequence No: 1, Text Type: D, B5


[9493050] Olympus followed up with the user facility via telephone and in writing to obtain additional information regarding the report, but received no additional information regarding the reported event. The device referenced in this report was returned to olympus for evaluation. The evaluation was able to duplicate the user's report of image difficulty. The image was intermittently distorted, and then lost. The evaluation also noted the shaft of the telescope was bent, dented, and scratched on the outer tube, and the unit did not pass light transmission testing. There was also burn marks and corrosion on the contact pins of the video connector. The cause of the failure was likely physical damage to the cable unit related to user handling. The device was serviced and returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2011-00035
MDR Report Key2352403
Report Source06
Date Received2011-11-18
Date of Report2011-11-22
Date of Event2011-11-21
Date Mfgr Received2011-10-21
Date Added to Maude2012-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2011-11-18
Returned To Mfg2011-10-21
Model NumberA50021A
Catalog NumberA50021A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-11-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.