MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-11-18 for OLYMPUS VIDEO TELESCOPE A50021A manufactured by Olympus Winter & Ibe Gmbh.
[2274282]
The user facility reported that during an unspecified procedure, the users experienced a complete loss of image. The procedure was completed using a different device. There was no further information provided.
Patient Sequence No: 1, Text Type: D, B5
[9493050]
Olympus followed up with the user facility via telephone and in writing to obtain additional information regarding the report, but received no additional information regarding the reported event. The device referenced in this report was returned to olympus for evaluation. The evaluation was able to duplicate the user's report of image difficulty. The image was intermittently distorted, and then lost. The evaluation also noted the shaft of the telescope was bent, dented, and scratched on the outer tube, and the unit did not pass light transmission testing. There was also burn marks and corrosion on the contact pins of the video connector. The cause of the failure was likely physical damage to the cable unit related to user handling. The device was serviced and returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9610773-2011-00035 |
MDR Report Key | 2352403 |
Report Source | 06 |
Date Received | 2011-11-18 |
Date of Report | 2011-11-22 |
Date of Event | 2011-11-21 |
Date Mfgr Received | 2011-10-21 |
Date Added to Maude | 2012-03-16 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | LAURA STORMS-TYLER |
Manufacturer Street | 2400 RINGWOOD AVENUE |
Manufacturer City | SAN JOSE CA 95131 |
Manufacturer Country | US |
Manufacturer Postal | 95131 |
Manufacturer Phone | 4848965688 |
Manufacturer G1 | OLYMPUS WINTER & IBE GMBH |
Manufacturer Street | KUEHNSTRASSE 61 |
Manufacturer City | HAMBURG 22045 |
Manufacturer Country | GM |
Manufacturer Postal Code | 22045 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | OLYMPUS VIDEO TELESCOPE |
Generic Name | VIDEO TELESCOPE |
Product Code | FBP |
Date Received | 2011-11-18 |
Returned To Mfg | 2011-10-21 |
Model Number | A50021A |
Catalog Number | A50021A |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | OLYMPUS WINTER & IBE GMBH |
Manufacturer Address | KUEHNSTRASSE 61 HAMBURG 22045 GM 22045 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2011-11-18 |