IMMAGE IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2011-12-04 for IMMAGE IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[2274206] The customer reported to the customer technical support (cts) that immage 800 instrument intermittently drops liquid on cover between sample probe wash well and sample wheel. The customer verified the tightness of all connection including draw tube, syringes and probe; no issues were noted. Patient results were not impacted by this event and no erroneous results were generated. The customer did not come in contact with the fluid. On (b)(6) 2011, a field service engineer (fse) performed the following: replaced the sample probe and performed all probe alignments. Ordered a new xcaliber 180? T-valve for the customer. The customer ran qc and no issues were noted by the fse
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2050012-2011-08229
MDR Report Key2357170
Report Source01,06
Date Received2011-12-04
Date of Report2011-11-03
Date of Event2011-11-03
Date Mfgr Received2011-11-03
Device Manufacturer Date2010-11-17
Date Added to Maude2012-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS DUNG NGUYEN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-12-04
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberN/A
ID NumberSW 2.0.10
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-04

© 2026 FDA.report