IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-12-06 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[2463444] Customer called on (b)(6) 2011 reporting of a fluid leak inside the immage 800 immunochemistry system. Customer stated that dried wash buffer was also found near the sample probe and under the instrument. Customer technical support (cts) advised customer the use of proper personal protective equipment before troubleshooting and instructed customer to check for active leaks at the probes and syringes but none were found. Cts also instructed customer to prime and check for further leaks and customer found pooled liquid under the reagent syringe assembly. No exposure or injury was reported as a result of the leak and patient results were not affected. Field service engineer (fse) spoke with the customer on the phone and the customer stated that they found 3 way valve fitting loose causing wash buffer to pool. Fse had the customer to remove the front cover and tightened fittings on valve assembly.
Patient Sequence No: 1, Text Type: D, B5


[9568873] Evaluation - method: troubleshooting was done over the phone. Evaluation - results: customer removed the front cover and tightened fittings on valve assembly. (b)(4). Troubleshooting was done over the phone.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-08252
MDR Report Key2360646
Report Source05,06
Date Received2011-12-06
Date of Report2011-11-14
Date of Event2011-11-14
Date Mfgr Received2011-11-14
Device Manufacturer Date2007-10-16
Date Added to Maude2012-07-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-12-06
Model NumberIMMAGE 800
Catalog NumberA15445
ID NumberSW VERSION 2.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.