HUTIV FINAL ASSY, STD, LEFT HUT ES 400002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2011-11-22 for HUTIV FINAL ASSY, STD, LEFT HUT ES 400002 manufactured by Liebel-flarsheim Co..

Event Text Entries

[2507134] On (b)(6): customer reports that during an undetermined urology procedure the sys fluoro failed. Staff moved the pt to a back up room where procedure was completed without further incident. Customer did not provide any pt or procedural info other than to say the procedure was completed, no reported injury.
Patient Sequence No: 1, Text Type: D, B5


[9493352] Covidien technical support began troubleshooting with biomed thinking the probable issue was the cpu pcb board. After trying several things over days of calls between biomed and tech support, including switching out the cpu pcb from another old unit in house, it was determined the actual problem was a blown pico fuse (3 amp f1) on the harness distribution board. There are no parts available for this sys. Tech support f/u; biomed reported that he replaced the 3 amp pico fuse himself and now the table is up and fully functional.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1518293-2011-00228
MDR Report Key2363092
Report Source06,07
Date Received2011-11-22
Date of Report2011-11-01
Date of Event2011-11-01
Date Mfgr Received2011-11-01
Device Manufacturer Date1993-05-01
Date Added to Maude2012-07-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID BENSON
Manufacturer Street2111 EAST GALBRAITH RD.
Manufacturer CityCINCINNATI OH 45237
Manufacturer CountryUS
Manufacturer Postal45237
Manufacturer Phone5139485719
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUTIV FINAL ASSY, STD, LEFT
Generic NameUROLOGY SUITE
Product CodeKQS
Date Received2011-11-22
Model NumberHUT ES
Catalog Number400002
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIEBEL-FLARSHEIM CO.
Manufacturer Address2111 EAST GALBRAITH RD. CINCINNATI OH 45237 US 45237


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-22