IMMULITE 2500 INSULIN N/A 030007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-12-09 for IMMULITE 2500 INSULIN N/A 030007 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[2447211] Note: date of event is approximately 2-3 years ago. Falsely elevated immulite 2500 insulin results were obtained on an immulite 2500 system for one patient. This patient was put on metformin.
Patient Sequence No: 1, Text Type: D, B5


[9446867] The physician for the patient that was put on metformin and incorrectly diagnosed as insulin sensitive, had been informed about the possibility of heterophilic antibody interference, it is believed that the patient was taken off metformin once the antibody interference was made known to the physician. As per the immulite 2500 insulin ifu, "heterophilic antibodies in human serum can react with the immunoglobulins included in the assay components causing interference with invitro immunoassays. Samples from patients routinely exposed to animals or animal serum products can demonstrate this type of interference potentially causing an anomalous result. These reagents have been formulated to minimize the risk of interference; however, potential interactions between rare sera and test components can occur. For diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings. "
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2011-00194
MDR Report Key2364277
Report Source05,06
Date Received2011-12-09
Date of Report2011-11-22
Date Mfgr Received2011-11-22
Date Added to Maude2012-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLINDA BARLETTA
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242299
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Street62 FLANDERS BARTLEY ROAD
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2500 INSULIN
Generic NameINSULIN
Product CodeCFP
Date Received2011-12-09
Model NumberN/A
Catalog Number030007
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer AddressGLYN RHONWY LLANBERIS CAERNAFON, GWYNEDD, WALES LL554EL UK LL55 4EL,


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-09

© 2024 FDA.report