* HT 1000R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-11-15 for * HT 1000R manufactured by Thermasolutions.

Event Text Entries

[2443407] While providing hyperthermic intraperitoneal chemotherapy (hipec), the thermachem machine had an electrical failure. After restarting the machine, i was able to resume the treatment after approximately thirty seconds. At this time medical engineering was paged to the room. After approximately a minute and a half the machine failed again. I restarted the machine again and continued with the treatment. Medical engineering had arrived and removed the side panel of the machine and while on the phone with a thermasolutions representative attempted to do a running repair of the machine. While the medical engineering personnel was doing the repair, the machine failed for a third time and i repeated what i had done before. After the machine was restarted, the medical engineer finished with the running repair and the machine had no further shutdowns. The medical engineer remained in the room for the remainder of the treatment delivery. Although i do not believe we were able to deliver an optimum treatment, it was near optimum and there was no harm done to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2366914
MDR Report Key2366914
Date Received2011-11-15
Date of Report2011-11-15
Date of Event2011-11-10
Report Date2011-11-15
Date Reported to FDA2011-11-15
Date Added to Maude2011-12-12
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameTHERMOCHEM HT1000R
Product CodeMLW
Date Received2011-11-15
Model NumberHT 1000R
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerTHERMASOLUTIONS
Manufacturer Address1889 BUERKLE ROAD BIRCHWOOD MN 55110 US 55110


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-15

© 2026 FDA.report