VP 2000 PROCESSOR 02J11-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-05-20 for VP 2000 PROCESSOR 02J11-001 manufactured by Abbott Molecular, Inc..

Event Text Entries

[2457085] The abbott vp 2000 processor is a device designed to automate and standardize slide specimen processing and routine slide staining for the laboratory. Customer reported having a leaking basin. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[9576226] Abbott molecular inc. Is aware of the problem with the vp 2000 processor basins. An urgent product recall (z-2313-2010) was distributed via fedex shipment to united states customers based on customer receipt of vp 2000 processor and basin spare part since september 2008. The recall was concluded in the united states in october, 2010; all leaking basins were replaced and all non-leaking basins were inspected by abbott field service personnel. Based on an additional complaint (see mdr 3005248192-2010-00012 for more information) another urgent product recall was issued fa cam-dec2010-093. This recall will coordinate replacement of all vp 2000 heater basins impacted by this manufacturing issue. The recall # z-2313-2010 was amended to include these additional replacements. The following mdrs were also filed: (b)(4). Three potential hazards were identified: delay in results. There is no harm associated with this hazard. Chemical exposure. There is no harm associated with this hazard as this issue does not increase the already expected exposure to these reagents. Therefore, there is no introduced operator harm. Slip hazard. Evaluation determined that the overall health risk was low, based on the following: the post correction leaking basing was 0. 4%. Leaking rate is expected to be about one to two drops per minute < 100 ul per min. ) when a loaded basin is either used in the vp2000 or stored on the lab bench, in the typical usage time < 8 hours), the leaked liquid would be less than 50 ml. Slip hazard would be avoided by the operator's observance of the liquid on the floor and by the small volume of liquid. If a leak were to occur, it would occur in an area that a lab operator would not normally walk (in the vp2000) or during reagent addition (lab bench). In addition, the leak would be visually obvious to the operator. If the operator did step on the leak, it would not necessarily lead to a fall that would lead to a serious injury.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005248192-2011-00008
MDR Report Key2370307
Report Source07
Date Received2011-05-20
Date of Report2011-05-23
Date of Event2011-05-03
Date Mfgr Received2011-05-03
Date Added to Maude2012-04-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR DIMITRIS DEMIRTZOGLOU
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617975
Manufacturer G1ABBOTT MOLECULAR, INC.
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal Code60018
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberZ-2313-2010
Event Type3
Type of Report3

Device Details

Brand NameVP 2000 PROCESSOR
Generic NameSTAINER, TISSUE, AUTOMATED, PRODUCT CODE: KEY
Product CodeKEY
Date Received2011-05-20
Catalog Number02J11-001
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.