[21752195] The abbott m2000 system is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the abbott m2000sp and the abbott m2000rt instruments. The abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing. The abbott m2000rt is an automated system for performing fluorescence-based pcr to provide quantitative and qualitative detection of nucleic acid sequences. A customer reported that the heated cover of the abbott m2000rt failed to achieve idle temperature in time. It smelled charred but no visible smoke or fire. No injury reported.
Patient Sequence No: 1, Text Type: D, B5


[21900417] This is likely a result from 24-volt heating circuit shortage located within the deep center of the m2000rt instrument, in a non-combustible environment. This shortage happens either during the instrument run or instrument initialization, during which the instrument door is closed and the heated cover is not accessible by the user. Similar to a burnt circuit, the event has no harm risk to the user (no electric shock hazard), and is not a fire hazard. The instrument was repaired.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005248192-2011-00003
MDR Report Key2370308
Report Source07
Date Received2011-04-29
Date of Report2011-04-29
Date of Event2011-04-04
Date Mfgr Received2011-04-04
Date Added to Maude2012-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR DIMITRIS DEMIRTZOGLOU
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617975
Manufacturer G1ABBOTT MOLECULAR, INC.
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal Code60018
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABBOTT M2000RT
Generic NameREAL TIME NUCLEIC ACID AMPLIFICATIONS SYSTEM, PRODUCT CODE: OOI
Product CodeOOI
Date Received2011-04-29
Catalog Number09K15-001
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-29