VP 2000 PROCESSOR 02J11-004

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-04-04 for VP 2000 PROCESSOR 02J11-004 manufactured by Abbott Molecular, Inc..

Event Text Entries

[2460552] The abbott vp 2000 processor is a device designed to automate and standardize slide specimen processing and routine slide staining for the laboratory. A customer and a field service representative in (b)(4) alleged that the metal edges of an abbott vp 2000 processor (heater block, covers) unit are very sharp. The parts with edges that are allegedly sharp include the cover for the ambient temperature baths and when the heated basins are removed, the openings of the heater reagent basins. Drying station and water bath have allegedly sharp edges also. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[9572342] Abbott molecular requested the return of the instrument with the alleged sharp edges for complaint investigation. The complaint is currently under investigation. Abbott molecular completed a sharpness test on a vp2000 instrument currently in use at abbot molecular. This instrument passed on all user accessible edges (per ul standard 1439).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005248192-2011-00002
MDR Report Key2370401
Report Source07
Date Received2011-04-04
Date of Report2011-04-04
Date of Event2011-03-07
Date Mfgr Received2011-03-07
Date Added to Maude2012-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR DIMITRIS DEMIRTZOGLOU
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617975
Manufacturer G1ABBOTT MOLECULAR, INC.
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal Code60018
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVP 2000 PROCESSOR
Generic NameSTAINER, TISSUE, AUTOMATED
Product CodeKEY
Date Received2011-04-04
Catalog Number02J11-004
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-04

© 2026 FDA.report