HYPERGLIDE OCCLUSION BALLOON SYSTEM 104-4112

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06 report with the FDA on 2011-10-12 for HYPERGLIDE OCCLUSION BALLOON SYSTEM 104-4112 manufactured by Ev3 Neurovascular.

Event Text Entries

[2502675] It was reported the balloon could not be inflated during procedure. It was reported that the balloon was tested for inflation/deflation during preparation and it had worked as intended. No pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[9572861] The balloon catheter was returned for evaluation with the guidewire stuck inside the catheter. The guidewire was found broken into two segments inside the catheter. Analysis of the cross section at the break point could not determined the cause of the guidewire failure. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2029214-2011-00332
MDR Report Key2371407
Report Source04,05,06
Date Received2011-10-12
Date of Report2011-09-30
Date of Event2011-08-22
Date Mfgr Received2011-09-30
Device Manufacturer Date2010-01-25
Date Added to Maude2012-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKIM SALCEDA
Manufacturer Street9775 TOLEDO WAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9498373700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHYPERGLIDE OCCLUSION BALLOON SYSTEM
Generic NameOCCLUSION BALLOON CATHETER
Product CodeNUF
Date Received2011-10-12
Returned To Mfg2011-09-13
Model Number104-4112
Lot Number9398916
Device Expiration Date2013-10-01
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerEV3 NEUROVASCULAR
Manufacturer Address9775 TOLEDO WAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2011-10-12

© 2024 FDA.report