IMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT N/A C1L2/C2L2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-12-16 for IMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT N/A C1L2/C2L2 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[19870266] Discordant false negative immulite 2000 xpi penicillin (c1 and c2) results were generated on one patient sample. The patient was first tested with a skin prick test with positive results. A sample was then tested on the immulite which gave negative results and when repeated they were slightly positive. The samples were then sent to a second laboratory where they tested positive. Several weeks later, a second sample was tested on the immulite 2000 xpi with negative results. Samples 1 and 2 were then sent to the swedish siemens allergy reference lab and both samples tested slightly positive. There was no known report of treatment given or withheld based on the discordant penicillin results.
Patient Sequence No: 1, Text Type: D, B5


[20280064] A siemens global product support engineer analyzed the immulite 2000 xpi instrument data. Analysis of the instrument data indicates that the cause for the discordant penicillin c1 and c2 results is unknown. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2011-00200
MDR Report Key2378471
Report Source05,06
Date Received2011-12-16
Date of Report2011-12-05
Date of Event2011-11-10
Date Mfgr Received2011-12-05
Date Added to Maude2012-07-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLINDA BARLETTA
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242299
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetGLYN RHONWY LLANBERIS
Manufacturer CityCAERNAFON, GWYNEDD,, WALES LL554EL
Manufacturer CountryUK
Manufacturer Postal CodeLL55 4EL
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT
Generic Name3GALLERGY SPECIFIC IGE
Product CodeDHB
Date Received2011-12-16
Model NumberN/A
Catalog NumberC1L2/C2L2
Lot Number344
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer AddressGLYN RHONWY LANNBERIS CAERNAFON, GWYNEDD,, WALES LL554EL UK LL55 4EL


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-16