HU-FRIEDY SUB-0 VISION CURETTE #6 SGCSUB-06

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-12-07 for HU-FRIEDY SUB-0 VISION CURETTE #6 SGCSUB-06 manufactured by Hu-friedy Mfg. Co., Llc.

Event Text Entries

[2367453] A periodontist was performing scaling and root planning on pt's supper left side between 12 & 13 when the curette broke off and became embedded in the pt's tissue. The doctor confirmed with an x-ray that the tip was embedded and then proceeded to perform a periodontal flap surgery for removal the tip and debridement of the area.
Patient Sequence No: 1, Text Type: D, B5


[9563363] Thickness at the heel of the unbroken working end measures 0. 0185" and the thickness in the middle of the blade measures 0. 0215". Both measurements are within the specification of 0. 020" +/- 0. 002. " based on visual/dimensional analysis, the reason for the breakage of the instrument cannot be determined. Add'l functional testing will be conducted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416605-2011-00004
MDR Report Key2379116
Report Source06
Date Received2011-12-07
Date of Report2011-12-05
Date of Event2011-10-17
Date Mfgr Received2011-11-21
Device Manufacturer Date2011-07-01
Date Added to Maude2012-07-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street3232 NORTH ROCKWELL ST.
Manufacturer CityCHICAGO IL 60618
Manufacturer CountryUS
Manufacturer Postal60618
Manufacturer Phone7739753975
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameHU-FRIEDY SUB-0 VISION CURETTE #6
Generic NameSGCSUB - 06
Product CodeEKT
Date Received2011-12-07
Returned To Mfg2011-11-21
Model NumberSGCSUB-06
Catalog NumberSGCSUB-06
Lot Number0711
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHU-FRIEDY MFG. CO., LLC
Manufacturer AddressCHICAGO IL 60618 US 60618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-12-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.