STAINLESS STEEL BLADE 08400402000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-11-07 for STAINLESS STEEL BLADE 08400402000 manufactured by Stryker Ireland Ltd..

Event Text Entries

[16195249] The cast cutter blade associated with this event was returned to the manufacturer for evaluation. It was visually confirmed that both the blade teeth and mount were worn. Lot number information has not been provided to permit further investigation. The ifu for the cast cutter blade has a caution which states "stryker does not recommend using sst blades ref 0840-040-200, 0840-040-250 and 0940-025-000 for cutting synthetic cast material. " the investigation results indicate that the user may have contributed to this event.
Patient Sequence No: 1, Text Type: N, H10


[16323608] It was reported that while removing a padded fiberglass cast, the patient received a cut on the arm. It was also reported that a minor injury was sustained, a break in the skin. It was further reported that the wound was dressed and no other treatment was required.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616696-2011-00216
MDR Report Key2379520
Report Source07
Date Received2011-11-07
Date of Report2011-10-07
Date of Event2011-10-07
Date Mfgr Received2011-10-07
Date Added to Maude2012-03-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactUNA BERRY
Manufacturer StreetCARRIGTWOHILL BUSINESS & TECHNOLOGY PARK
Manufacturer CityCARRIGTWOHILL, CORK
Manufacturer CountryEI
Manufacturer Phone214532963
Manufacturer G1STRYKER IRELAND LTD.
Manufacturer StreetCARRIGTWOHILL BUSINESS & TECHNOLOGY PARK
Manufacturer CityCARRIGTWOHILL, CORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTAINLESS STEEL BLADE
Product CodeLGH
Date Received2011-11-07
Returned To Mfg2011-10-20
Catalog Number08400402000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER IRELAND LTD.
Manufacturer AddressCARRIGTWOHILL, CORK EI


Patients

Patient NumberTreatmentOutcomeDate
10 2011-11-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.