WIRE FX BOLT SLOTTED SS 818050006

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2011-12-16 for WIRE FX BOLT SLOTTED SS 818050006 manufactured by Depuy Warsaw.

Event Text Entries

[2370350] The two pieces are stuck together.
Patient Sequence No: 1, Text Type: D, B5


[9563862] This complaint is still under investigation. Depuy will notify the fda of the results of this investigation once it has been completed.
Patient Sequence No: 1, Text Type: N, H10


[16926310] Examination of the returned products confirmed the complaint. The cause is attributed to the suppliers manufacturing process, the major diameter of the bolt threads is oversized up to. 004, preventing the nut from fastening to the bolt. The thread form is sharp at the crest, compared to conforming bolts. Hold order (b)(4) was initiated on (b)(4) 2011. An hhe was conducted and is documented on (b)(4), eco# (b)(4). The lot numbers affected were recalled, root causes and corrective actions are captured in capa (b)(4) that was initiated on (b)(4) 2011. Depuy considers the investigation closed. Should any additional information be received to change the outcome of the performed investigation, the complaint will be re-opened.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1818910-2011-25227
MDR Report Key2380075
Report Source05,08
Date Received2011-12-16
Date of Report2011-11-30
Date Mfgr Received2011-11-30
Date Added to Maude2011-12-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHERINE SEPPA
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer G1DEPUY WARSAW
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWIRE FX BOLT SLOTTED SS
Generic NameTRAUMA FIXATION
Product CodeHXK
Date Received2011-12-16
Catalog Number818050006
Lot Number582520
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY WARSAW
Manufacturer Address700 ORTHOPAEDIC DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-16

© 2024 FDA.report