IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-12-30 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2373459] The affiliate reported the customer alleged the sample probe wash station overflowed and did not drain involving immage 800 immunochemistry system. Fluid leaked onto the top of the unit. Stopping and resetting the system did not resolve the issue. Beckman coulter customer technical support (cts) advised the customer to turn the unit off and reboot. The customer stated a complete shutdown and reboot appeared to have resolved the issue. No patient samples were run prior to or during the event. No patient results were generated. There was no injury or adverse affect associated with this event. The unit was in operation.
Patient Sequence No: 1, Text Type: D, B5


[9680655] The customer contacted beckman coulter, inc. On (b)(4) 2011 and stated the same issue returned. Beckman coulter customer technical support (cts) assisted the customer to replace the sample valve over the telephone. The customer confirmed the unit was in operation with no further issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-08496
MDR Report Key2392794
Report Source01,05,06
Date Received2011-12-30
Date of Report2011-12-01
Date of Event2011-12-01
Date Mfgr Received2011-12-01
Device Manufacturer Date2009-11-11
Date Added to Maude2012-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-12-30
Model NumberNA
Catalog NumberA15445
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.