THERMOCHEM HT1000R HT-1000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-12-22 for THERMOCHEM HT1000R HT-1000 manufactured by Thermasolutions.

Event Text Entries

[2371012] Title: xxxxx. Event desc: sixty minutes into a ninety minute hyperthermic intraperitoneal chemoperfusion (hipec) procedure, the thermochem ht 1000r device stopped functioning. The device was on loan from the company thermasolutions. The following notice was on the video screen "no video input. " the video cable at the rear of the device was noticed to be defective and damaged. We called the company rep and took steps via phone instructions to bring the device back on line. The procedure was finished without further issues. Device usage problem: other.
Patient Sequence No: 1, Text Type: D, B5


[9682200] A service engineer went to the location after the case. The service engineer replaced the over temperature switch and then evaluated and adjusted voltage levels. The device was then run through an installation process to confirm the repairs corrected the issue reported by the customer. The device was operating properly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005699784-2011-00001
MDR Report Key2393952
Report Source07
Date Received2011-12-22
Date of Report2011-11-15
Date of Event2011-11-03
Report Date2011-11-01
Date Reported to FDA2011-11-01
Date Mfgr Received2011-11-29
Device Manufacturer Date2001-05-01
Date Added to Maude2012-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street1889 BUERKLE RD
Manufacturer CityWHITE BEAR LAKE MN 55110
Manufacturer CountryUS
Manufacturer Postal55110
Manufacturer Phone6512093900
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHERMOCHEM
Generic NameTHERMOCHEM HT 1000R
Product CodeMLW
Date Received2011-12-22
Returned To Mfg2011-11-17
Model NumberHT1000R
Catalog NumberHT-1000
Lot NumberNONE
ID NumberNONE
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTHERMASOLUTIONS
Manufacturer Address1889 BUERKLE RD BIRCHWOOD MN 55110 US 55110


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-12-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.