GCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN COMBILITE TYPE 4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01 report with the FDA on 2011-12-28 for GCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN COMBILITE TYPE 4 manufactured by .

Event Text Entries

[2399011] This is an initial spontaneous incident report found in a newspaper in the (b)(6) concerning a female pt, of unk age, hospitalised in an intensive care unit. The initial information on this incident was received on (b)(4) 2011 but the type of valve was not communicated to the company until on (b)(4) 2011 an oxygen cd cylinder was used for providing respiratory oxygen while transferring the pt. The oxygen cylinder was placed between the pt's legs on the bed for transfer. When the valve, a gce combilite i vipr, 40 microns (combilite type 4 (boc part number 19317500)) for oxygen, was turned on, blue flames were seen coming out of the outlet. This caused 18% burns (both legs) to the pt and burns on the bed and the floor. The pt was transferred to the burns unit at (b)(6) hospital. The current condition of the pt is unk. The burns were assessed as a medically significant event. The valve is with (b)(4) health and safety executive for investigations. Linde gas north america llc has a 510 (k) for a similar, but not identical, product. Fda us will be informed. Additional information is expected.
Patient Sequence No: 1, Text Type: D, B5


[9683282] Accidental spontaneous oxygen valve ignition causing severe burns requiring burn unit therapy. The valve is undergoing intensive and extensive investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003900188-2011-00030
MDR Report Key2396786
Report Source00,01
Date Received2011-12-28
Date Mfgr Received2011-11-29
Device Manufacturer Date2009-11-01
Date Added to Maude2012-01-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN
Generic NameNONE
Product CodeECX
Date Received2011-12-28
Model NumberCOMBILITE TYPE 4
Lot Number4411485
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-12-28