IMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT N/A ALLERGEN F10

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-01-09 for IMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT N/A ALLERGEN F10 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[2330596] False negative immulite 2000 f10 (sesame seed) allergen results were generated on three patient samples. The laboratory reran the samples on a second instrument for confirmation. All repeat results were positive. There was no known report of treatment given or withheld based on the discordant.
Patient Sequence No: 1, Text Type: D, B5


[9595705] A siemens technical support specialist analyzed the immulite 2000 instrument data. Analysis of the instrument data indicates that the cause for the discordant. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2012-00001
MDR Report Key2403594
Report Source05,06
Date Received2012-01-09
Date of Report2011-12-15
Date of Event2011-12-15
Date Mfgr Received2011-12-15
Date Added to Maude2012-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLINDA BARLETTA
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242299
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetGLYN RHONWY LLANBERIS
Manufacturer CityCAERNAFON, GWYNEDD,, WALES LL554EL
Manufacturer CountryUK
Manufacturer Postal CodeLL55 4EL
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2000 3GALLERGY SPECIFIC IGE UNIVERSAL KIT
Generic Name3GALLERGY SPECIFIC IGE
Product CodeDHB
Date Received2012-01-09
Model NumberN/A
Catalog NumberALLERGEN F10
Lot Number202
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer AddressGLYN RHONWY LANNBERIS CAERNAFON, GWYNEDD,, WALES LL554EL UK LL55 4EL


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.