TRAY,SKIN SCRUB,VAG,PVP-I 4469A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-01-12 for TRAY,SKIN SCRUB,VAG,PVP-I 4469A manufactured by Carefusion.

Event Text Entries

[2564789] The following event description was taken from a medwatch sent to us by the fda: sponges are coming off the sticks rom the scrub prep kits. " writer received a return phone call from (b)(6) at the facility regarding this event she indicated that there was no harm to the patient and the procedure continued without interruption. Writer asked about the sample return and she indicated that in most cases the samples do come to her for return to the manufacturer, but unfortunately in this cas,e she did not receive them to forward on to carefusion for evaluation.? She did indicate that there is a chance that the samples are with her cooperate purchaser but she can't verify this. At this time, no samples are available for evaluation. No sample was provided by the customer; therefore, an evaluation of the complaint device for deficiency of construction could not be performed. However, samples of our current production process were reviewed and no problems related with this issue were observed. However, in similar complaint, this incident was identified the contour stick sponges were without seal. Therefore, to confirm the issue reported on this complaint is necessary the sample or picture. The device history records for the lot reported were evaluated for any issues related with this customer report. The product was manufactured, inspected and released in accordance with our internal procedures and no issues were observed our manufacturing process was reviewed and we could observed that the work station for contour stick sponge, was re-designed where was separated from the work tables in order to maintain the pre-assembly of the contour stick sponges (unsealed) separated from the sealed stick sponges. In addition, the pre-assembly contour stick sponges (unsealed) are not accumulate in order to prevent that the sponges mixes with the stick sponges sealed. A tension test is performed during our manufacturing process in order to verify the seal on the stick sponges ( contour & square). In addition, samples are taken from production every 2 hours to conduct a destructive pull test to detect any anomalies in the welding process. The sample size is based on an aql. 65, which is a 95% ci and we accept with 0 defects and reject with 1.
Patient Sequence No: 1, Text Type: D, B5


[9582036] (b)(4). "originally the decision was made that this complaint did not meet the requirements for mdr submission in adherence with 21 code of federal regulation part 803. However, due to a similar event subsequently being submitted as an mdr, the decision has been made to retrospectively submit this event as an mdr. " an inquiry was received from the fda and sent previously, this document has been attached to this mdr.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030673-2012-00071
MDR Report Key2413451
Report Source05,06
Date Received2012-01-12
Date of Report2012-01-12
Date of Event2011-06-01
Date Mfgr Received2011-07-12
Date Added to Maude2012-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8474737294
Manufacturer G1PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V
Manufacturer StreetCERRADA VIA DE LA PRODUCCION NO85PARQUE INDUST.MEXICALI III
Manufacturer CityMEXICALI 21600
Manufacturer CountryMX
Manufacturer Postal Code21600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRAY,SKIN SCRUB,VAG,PVP-I
Generic NameAPPLICATOR, ABSORBENT TIPPED, STERILE
Product CodeOJU
Date Received2012-01-12
Model Number4469A
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address1500 WAUKEGAN ROAD MCGAW PARK IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-12