GPS III SINGLE KIT W/30ML ACDA N/A 800-1003A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-01-18 for GPS III SINGLE KIT W/30ML ACDA N/A 800-1003A manufactured by Biomet Biologics.

Event Text Entries

[2427589] It was reported that during a procedure on (b)(6) 2011, patient's blood was placed in a platelet concentration kit. As the blood was put into the cylinder, the blood started to be expelled out the top of the cylinder. Upon inspection of the cylinder, it was noted that there was a hole on the top of the cylinder where a filter is usually located.
Patient Sequence No: 1, Text Type: D, B5


[9535283] Current information is insufficient to permit a conclusion as to the cause of the event. Review of receiving certificate of conformance showed that lot released with no recorded anomaly or deviation. Review of device sales history shows eight devices confirmed used with no reported complications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0001825034-2012-00061
MDR Report Key2420597
Report Source07
Date Received2012-01-18
Date of Report2011-12-19
Date of Event2011-12-19
Date Mfgr Received2011-12-19
Device Manufacturer Date2011-10-07
Date Added to Maude2012-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. ANGELA DICKSON
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5742676639
Manufacturer G1BIOMET BIOLOGICS
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameGPS III SINGLE KIT W/30ML ACDA
Generic NameSUPPLIES, BLOOD BANK
Product CodeKSS
Date Received2012-01-18
Model NumberN/A
Catalog Number800-1003A
Lot Number112953
ID NumberN/A
Device Expiration Date2014-09-30
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMET BIOLOGICS
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.