VARI-STAT VARIABLE LOW TEMP CAUTERY 84-40000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 1999-09-30 for VARI-STAT VARIABLE LOW TEMP CAUTERY 84-40000 manufactured by Xomed Surgical Products, Inc..

Event Text Entries

[135754] It was reported by the user facility, "the cautery was used to mark the location of a tear on the sclerotic in a detachment of the retina. It did not function as usual. At the same level on the rheostat, it overheated which lead to the incident. The outcome of the incident was a perforation of the sclerotic of the chroid and leaking of the vitreous body. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-1999-00009
MDR Report Key242799
Report Source01,05
Date Received1999-09-30
Date of Report1999-08-31
Date of Event1999-01-01
Date Mfgr Received1999-08-31
Device Manufacturer Date1999-03-01
Date Added to Maude1999-10-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMARK HIMES
Manufacturer Street6743 SOUTHPOINT DR NORTH
Manufacturer CityJACHSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9042797578
Manufacturer G1XOMED SURGICAL PRODUCTS, INC.
Manufacturer Street6743 SOUTHPOINT DR NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVARI-STAT VARIABLE LOW TEMP CAUTERY
Generic NameCAUTERY
Product CodeHQP
Date Received1999-09-30
Model NumberNA
Catalog Number84-40000
Lot Number17838700
ID NumberNA
Device Expiration Date2003-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key235227
ManufacturerXOMED SURGICAL PRODUCTS, INC.
Manufacturer Address6743 SOUTHPOINT DR. NORTH JACKSONVILLE FL 32216 US
Baseline Brand NameVARI-STAT VARIABLE LOW TEMP CAUTERY
Baseline Generic NameCAUTERY
Baseline Model NoNA
Baseline Catalog No84-40000
Baseline IDNA
Baseline Device FamilyCAUTERY
Baseline Shelf Life ContainedY
Baseline Shelf Life [Months]48
Baseline PMA FlagN
Baseline 510K PMNN
Baseline PreamendmentY
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1999-09-30